At a glance
- Biological evaluation under ISO 10993-1:2025 concerns the medical device in its final state, not the granules.
- Under Regulation (EU) 2017/745, the manufacturer of the medical device is responsible.
- The term USP Class VI comes from a pharmacopoeia chapter on packaging whose revision drops the plastic classes.
- The VDI 2017 guideline, edition August 2026, describes formulation consistency, security of supply and change management.
- Steam, ethylene oxide, radiation and hydrogen peroxide stress plastics differently, so the sterilisation method belongs in every enquiry.
Biocompatibility: what is evaluated and by whom
Who evaluates biocompatibility?
Biocompatibility is evaluated by the manufacturer of the medical device, on the product in its final state. Regulation (EU) 2017/745 obliges the manufacturer to ensure that its devices meet the requirements. Annex I requires attention to toxicity and to compatibility with tissues and body fluids when choosing materials.[5]
The framework of the evaluation is set by ISO 10993-1. The standard has been available in its sixth edition since November 2025, restructured and aligned with risk management under ISO 14971.[1] The US FDA's guidance on the standard dates from 2023, before the new edition. It considers the device in its final form, including processing and sterilisation. Material data or supplier information alone do not demonstrate the biocompatibility of the finished device, because processing can change the chemistry of the part.[4]
Which parts of ISO 10993 concern the plastic?
For the material, the parts on cytotoxicity, sensitisation, irritation and chemical characterisation matter most. Part 5 tests cytotoxicity on cell cultures, on the device or on extracts.[2]
| Document and edition | Subject | Relevance for the plastic |
|---|---|---|
| ISO 10993-1:2025 | Principles of biological evaluation within risk management | framework; the medical device is evaluated |
| ISO 10993-5:2009 | Tests for in vitro cytotoxicity with cell cultures | test on the device or on extracts |
| ISO 10993-10:2021 | Tests for skin sensitisation | applies to devices and their materials |
| ISO 10993-23:2021 with Amendment 1:2025 | Tests for irritation | amendment adds in vitro skin models |
| ISO 10993-18:2020 with Amendment 1:2022 | Chemical characterisation of materials | extractables and leachables; also addressed to material suppliers |
| VDI 2017:2026-08 | Medical Grade Plastics | formulation consistency, security of supply, change management |
The texts of the standards are subject to a charge and were not available to the editors. The information comes from the public abstracts of the issuing bodies. Regulation (EU) 2017/745 was read in its freely available full text.
What does USP Class VI mean?
USP Class VI is a plastic class from Chapter 88 of the US Pharmacopeia and not an approval for medical devices. The chapter covers plastics, elastomers and other polymers in drug packaging and combination products.[7] The revision drops the table of plastic classes; the systemic injection test remains as an animal test. Chapter 87, with in vitro tests, is extended. According to the professional society ISPE, both revisions apply from 1 December 2026.[7][8]
Sterilisation: methods, standards and effects on the material
Which sterilisation methods are there?
Steam, ethylene oxide, ionising radiation and vaporised hydrogen peroxide are common. For each method an ISO standard governs development, validation and routine control of the process.[9][10][11] Regulation (EU) 2017/745 requires validated methods for devices supplied sterile.[5]
| Method | Standard and edition | Conditions according to the literature | Effect on polymers |
|---|---|---|---|
| Moist heat (steam) | ISO 17665:2024 | about 120 °C at 103 to 117 kPa | hydrolysis of sensitive polymers |
| Ethylene oxide | ISO 11135:2014 with Amendment 1:2018 | 37 to 63 °C, 1 to 6 hours | reaction with amine, carboxyl, thiol and hydroxyl groups possible |
| Gamma radiation | ISO 11137-1:2025 | 15 to 45 kGy, on average about 25 kGy | chain scission and crosslinking |
| Electron beam | ISO 11137-1:2025 | typically 25 kGy | as gamma radiation, lower penetration depth |
| Vaporised hydrogen peroxide | ISO 22441:2022 | 25 to 50 °C, as plasma 40 to 65 °C | oxidative crosslinking by hydroxyl radicals |
How does sterilisation change the plastic?
Each method attacks the plastic in its own way: steam through heat and moisture, radiation and hydrogen peroxide through radicals. According to a review on polymer implants, steam can degrade polymers sensitive to hydrolysis. Gamma and electron beams cut chains and crosslink them, changing molar mass, crystallinity and mechanical behaviour.[12]
For ultra-high-molecular-weight polyethylene, the same review reports embrittlement after hydrogen peroxide plasma similar to that after gamma irradiation.[12] These are research examples on implant materials, not an overview of all engineering plastics. Under ISO 10993-18, residues from sterilisation are among the substances to be considered in chemical characterisation.[3]
Change control: formulation consistency and medical grade
What does the VDI 2017 guideline govern?
The VDI 2017 guideline of the Association of German Engineers defines what is meant by medical grade plastics and which requirements apply to them. The current edition dates from August 2026 and replaces the edition of July 2019. Before that, the term medical grade was not precisely defined in standardisation.[6]
The guideline covers formulation consistency, security of supply, change management, and production, packaging, storage and logistics. It adds information, documentation and testing obligations between supplier and customer, and a checklist.[6] The public abstract does not state specific deadlines for change notifications.
Why is a formulation change critical?
A changed formulation can call the biological evaluation of the medical device into question. The FDA names a change of plastic supplier as an example of a change that can affect biocompatibility.[4] Under Regulation (EU) 2017/745, the manufacturer must maintain a risk management system.[5] How change notifications are agreed is described in VDI 2017.[6]
Applications and limits
Where are plastics used in medical and hygiene applications?
Plastics are found in devices with every kind of body contact, from the skin surface to the implant.[1]
- Surface contact: devices on the skin, plus protective equipment such as gloves and surgical masks.[1]
- Externally communicating devices: ISO 10993-1 names them as a separate group.[1]
- Implants: the technical literature names, for example, PEEK for spinal cages.[12]
- In vitro diagnostics and pharmaceutical packaging: VDI 2017 includes both areas.[6]
What does a medical grade plastic not do?
A medical grade plastic does not replace the evaluation of the medical device.
- No approval: test reports on the material are a starting point. Responsibility remains with the manufacturer of the medical device.[4][5]
- Processing counts: testing is done on a sample that has gone through the same manufacturing and sterilisation steps as the device.[4]
- Substance restrictions: Annex I of the Regulation requires a justification if certain devices contain carcinogenic, mutagenic or toxic to reproduction substances above 0.1 % by mass.[5]
What the material supplier needs to know
With this information the material supplier can name the appropriate evidence:
- Type of body contact: skin, mucosa, blood, tissue or no contact
- Duration of contact and risk class of the medical device
- Sterilisation method, number of cycles and, for radiation, the dose
- Contact with disinfectants, drugs or body fluids
- Required evidence on the material, for example tests to ISO 10993-5
- Formulation consistency, notice period for change notifications and supply period
- Colour and requirements for additives
- Mechanical requirements after sterilisation
- Target markets, quantities and processing method
Frequently asked questions
Are there approved plastics for medical devices?
No. The medical device is evaluated, not the raw material. Test data on the plastic are building blocks for the evaluation by the device manufacturer.
What does medical grade mean for plastics?
The VDI 2017 guideline defines what is meant by medical grade plastics. Among other things it covers formulation consistency, security of supply and change management.
Which edition of ISO 10993-1 is current?
The current edition is ISO 10993-1:2025, the sixth edition of November 2025.
Can a polyamide be tested to ISO 10993?
Individual tests, for example for cytotoxicity to ISO 10993-5, are possible on the material or on extracts. The biological evaluation under ISO 10993-1, however, concerns the finished medical device.
Is USP Class VI an approval for medical devices?
No. The term comes from Chapter 88 of the US Pharmacopeia on polymers in drug packaging. The revision drops the plastic classes.
Who is responsible under Regulation (EU) 2017/745?
The manufacturer is responsible: whoever manufactures a medical device or has it manufactured and markets it under its own name or trademark.
Which standards apply to the sterilisation of medical devices?
ISO 17665:2024 covers moist heat, ISO 11135:2014 ethylene oxide, ISO 11137-1:2025 radiation and ISO 22441:2022 vaporised hydrogen peroxide.
How does gamma sterilisation affect plastics?
Gamma radiation creates radicals in the polymer that cut and crosslink chains. The technical literature names doses of 15 to 45 kGy.
Why must the material supplier report changes?
A changed formulation can call the biological evaluation of the medical device into question. The FDA names a change of plastic supplier as an example.
Sources
All sources retrieved and checked on 23 November 2026.
- ISO 10993-1:2025, Biological evaluation of medical devices – Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process. 6th edition, November 2025. Abstract. https://www.iso.org/standard/10993-1
- ISO 10993-5:2009, Part 5: Tests for in vitro cytotoxicity. 3rd edition, June 2009, confirmed 2022. Abstract. https://www.iso.org/standard/36406.html
- ISO 10993-18:2020 with Amendment 1:2022, Part 18: Chemical characterization of medical device materials within a risk management process. 2nd edition, January 2020, confirmed 2025. Abstract. https://www.iso.org/standard/64750.html
- U.S. Food and Drug Administration: Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1. Guidance for Industry and FDA Staff, 8 September 2023. https://www.fda.gov/media/142959/download
- Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices. Article 2(30), Article 10, Annex I sections 10 and 11. https://eur-lex.europa.eu/legal-content/EN/TXT/HTML/?uri=CELEX:32017R0745
- VDI 2017:2026-08, Medical Grade Plastics (MGP). Edition August 2026, replaces VDI 2017:2019-07. Abstract and table of contents. https://www.dinmedia.de/de/technische-regel/vdi-2017/403537122
- United States Pharmacopeia: General Chapter 88, Biological Reactivity Tests, In Vivo. Public preview of the chapter. https://doi.usp.org/USPNF/USPNF_M98834_02_01.html
- International Society for Pharmaceutical Engineering (ISPE): article on the revisions of USP Chapters 87 and 88 and the removal of the plastic classes. https://ispe.org/pharmaceutical-engineering/ispeak/learn-about-future-usp-plastics-biotech
- ISO 17665:2024, Sterilization of health care products – Moist heat. 1st edition, March 2024. Abstract. https://www.iso.org/standard/80271.html
- ISO 11135:2014 with Amendment 1:2018, Sterilization of health-care products – Ethylene oxide. 2nd edition, July 2014, confirmed 2024. Abstract. https://www.iso.org/standard/56137.html
- ISO 11137-1:2025, Sterilization of health care products – Radiation – Part 1. 2nd edition, April 2025. Abstract. https://www.iso.org/standard/81721.html
- Herczeg, C. K.; Song, J.: Sterilization of Polymeric Implants: Challenges and Opportunities. ACS Applied Bio Materials 5 (2022), pp. 5077–5088. https://doi.org/10.1021/acsabm.2c00793
Related topics
- Food contact
Regulation (EU) No 10/2011, migration and the declaration of compliance for plastics in contact with food. - Drinking water contact
Rules and approvals for plastics in contact with drinking water, in Germany, Europe and other countries. - PPS and PEEK
Chemical resistance, fire behaviour and continuous use temperature. - Colour
Colour matching, colour systems, colour difference and laser marking.
This page is technical information, not advice on standards, law or design. The standards cited in their current edition and the material manufacturers' test reports are authoritative.